ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Event Description
The initial reporter stated they received questionable results for one patient sample tested with the elecsys ft4 iii assay on a cobas 6000 e 601 module and a cobas 8000 e 602 module.The results measured at the customer site were reported outside of the laboratory to a physician.Values highlighted in yellow are questionable.The sample was initially measured on the customer's e 601 analyzer on (b)(6) 2021.The sample was repeated on an abbott architect analyzer.The sample was also provided for investigation, where it was tested on an e 602 analyzer on (b)(6) 2021.The serial number of the customer's e 601 analyzer is (b)(4).Ft4 reagent lot number 54716805, with an expiration date of 31-may-2022 was used on this analyzer.The serial number of the e 602 analyzer used for investigation is (b)(4).Ft4 reagent lot number 547168, with an expiration date of 31-may-2022 was used on this analyzer.
 
Manufacturer Narrative
Phone number was provided as (b)(6).Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods and the standardization methodology used.The investigation could not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13051479
MDR Text Key288632304
Report Number1823260-2021-03803
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot Number54716805
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2021
Initial Date FDA Received12/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age16 YR
Patient SexFemale
-
-