• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE 3.8C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE 3.8C Back to Search Results
Model Number EG-3870UTK
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2021
Event Type  malfunction  
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Manufacturer Narrative
We checked the returned unit and confirmed the ccd driver pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd driver pcb.In addition, we confirmed that the operation channel buckled and the objective gluing missing; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
LINEAR ULTRASOUND VIDEO GASTROSCOPE 3.8C
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-10
akishima-shi
tokyo, 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-10
akishima-shi
tokyo, 196-0 012
JA   196-0012
Manufacturer Contact
william
3 paragon drive
montvale
nj 07645
MDR Report Key13058046
MDR Text Key287196755
Report Number9610877-2021-50920
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K182004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG-3870UTK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2021
Initial Date FDA Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-