• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG E-LUMINEXX VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG E-LUMINEXX VASCULAR STENT Back to Search Results
Catalog Number ZVM10100
Device Problems Misfire (2532); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2021
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the e-luminexx vascular stent that are cleared in the us.The pro code and 510 k number for the e-luminexx vascular stent are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending and two photos were provided for review.The investigation of the reported event is currently underway.Expiration date: 12/2024.
 
Event Description
It was reported that during a stent placement procedure, the device allegedly partially deployed.It was further reported that had to open it up from the handle & deployed using pin & pull method.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the e-luminexx vascular stent that are cleared in the us.The pro code and 510 k number for the e-luminexx vascular stent are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was returned for evaluation.Provided photos show part of the delivery system detached from the handgrip.During evaluation, the handgrip was found completely disassembled and the stent was not found in the catheter as it was already deployed.Based on evaluation of the provided photos and returned sample, the investigation is confirmed for detachment.A definite root cause of the reported incident can not be identified.Labeling review: in reviewing the relevant labeling it was found that the instructions for use sufficiently address the potential risks.With regards to general warnings, the instructions for use states that "should unusual resistance be felt at any time during the procedure, the entire system (introducer sheath or guiding catheter and stent delivery system) should be removed as a single unit".Regarding warnings, the instructions for use states "visually inspect the bard e-luminexx vascular stent to verify that the device has not been damaged due to shipping or improper storage.Do not use damaged equipment".H10: d4 (expiry date: 03/2024), g3, h6 (device).H11: h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that during a stent placement procedure, the device allegedly partially deployed.It was further reported that had to open it up from the handle & deployed using pin & pull method.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E-LUMINEXX VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13058476
MDR Text Key282964207
Report Number9681442-2021-00635
Device Sequence Number1
Product Code NIO
UDI-Device Identifier00801741147166
UDI-Public(01)00801741147166
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P080007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberZVM10100
Device Lot NumberANFQ2175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2021
Initial Date FDA Received12/22/2021
Supplement Dates Manufacturer Received01/31/2022
Supplement Dates FDA Received02/02/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-