Model Number 9735602
Device Problems Material Integrity Problem (2978); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9660651r, serial/lot #: (b)(4), ubd: unknown, udi#: unknown, device evaluated by mfr: no products have been returned to medtronic for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that after receiving the system back from planned maintenance (pm), an error message, "8" was showing when booting up the breakout box.The navigation light on the box was not illuminating green.The manufacturer representative (rep) stated that the system was working fine for the last year prior to pm.The rep attempted troubleshooting, but the light/breakout box never came on.There was no patient involvement.
 
Manufacturer Narrative
H3) the axiem was returned for analysis.Functional testing determined that the component was functioning as designed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13059628
MDR Text Key282613014
Report Number1723170-2021-02938
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/20/2021
Initial Date FDA Received12/22/2021
Supplement Dates Manufacturer Received01/27/2022
Supplement Dates FDA Received02/03/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
"SEE H10...."
-
-