• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® CPT¿ NH FICOLL¿; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® CPT¿ NH FICOLL¿; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 362780
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd had not received samples, but one (1) photo was provided for investigation.The photo was reviewed and the indicated failure mode for glass breakage during centrifugation was observed, as the photo shows broken tubes in the centrifuge.Additionally, twenty-three (23) retention samples from bd inventory were evaluated by visual examination and the issue of glass breakage was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality notifications.This complaint has been confirmed for the indicated failure mode of glass breakage during centrifugation based on the photo provided.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd vacutainer® cpt¿ nh ficoll¿ glass tube broke during the centrifugation.The following information was provided by the initial reporter: "tubes centrifuged at 1800g for 30min following ifu.Sometimes during centrifugation, the tubes break"."tube breakage during centrifugation incident happened: during use".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® CPT¿ NH FICOLL¿
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13061378
MDR Text Key285014393
Report Number1917413-2021-01079
Device Sequence Number1
Product Code JCF
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K891407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2021
Device Catalogue Number362780
Device Lot Number0311203
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-