(b)(4) primary rn initiated avvh (accelerated venovenous hemofiltration) treatment for patient.Shortly after initiation, rn noticed replacement fluid dripping from the nxstage machine.Per rn, fluid was dripping very fast.Blood returned to patient and rn removed cartridge and inspected it.Replacement fluid appeared to be leaking from balance chamber area.Educator notified, nxstage representative has been contacted about the issue.Defective cartridge was not saved.Car 500 crrt cartridge for nxstage system one.(b)(4) - rn went to start cvvh (continuous venovenous hemofiltration) and the tubing on the cartridge was crimped and were unable to use this cartridge (b)(4), lot# 10778017.(b)(4) - patient on continuous dialysis (cvvh) - cvvh machine with frequent therapy fluid inlet air detected alarms that started overnight.Upon closer inspection this morning tubing on fluid warmer kinked on several of the connectors to the therapy fluid bags.Bladder on fluid warmer frequently empty and difficult to fill.Opened 2 new sets of tubing that was left over from other cartridges (same lot number) and several of the connectors noted to be very kinked as well.One set of connectors without any kinks found out of 6 (in the 2 new bags) and connected to therapy fluid bags and warmer (no alarms since).Pictures of tubing taken and also sent to umass nxstage representative.She's filling out their internal incident report and is sending box to return tubing to them (awaiting incident report number) (b)(4), lot# 10878016.(b)(4) - nurse reported that the pressures rapidly increased and then the clamp on the filter unclamped on its own, the syringe shot off, and subsequently bled onto the floor.(b)(4).
|