• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRA ICE PLUS; CATHETER, ULTRASOUND, INTRACARDIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRA ICE PLUS; CATHETER, ULTRASOUND, INTRACARDIAC Back to Search Results
Model Number 86200
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
During an ablation to treat left atrial flutter a ultra ice plus catheter and rhythmia mapping system were selected for use.It was reported that poor ultrasound imaging/noise was observed during the procedure.The catheter was replaced, however, the transeptal puncture was unsuccessful.The procedure was aborted.Further troubleshooting indicated that the cause of the noise was the rhythmia signal station.
 
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA ICE PLUS
Type of Device
CATHETER, ULTRASOUND, INTRACARDIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key13078452
MDR Text Key282725573
Report Number2134265-2021-16174
Device Sequence Number1
Product Code ITX
UDI-Device Identifier08714729904380
UDI-Public08714729904380
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/30/2023
Device Model Number86200
Device Catalogue Number86200
Device Lot Number0027756703
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
RHYTHMIA HDX MAPPING SYSTEM
-
-