Model Number 87038
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
During an ablation to treat left atrial flutter a ultra ice plus catheter and rhythmia mapping system were selected for use.It was reported that poor ultrasound imaging/noise was observed during the procedure.The catheter was replaced, however, the transeptal puncture was unsuccessful.The procedure was aborted.Further troubleshooting indicated that the cause of the noise was the rhythmia signal station.
 
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHYTHMIA HDX
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
EP TECHNOLOGIES, INC.
4100 hamline avenue north
saint paul MN 55112
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key13078456
MDR Text Key282725827
Report Number2134265-2021-16175
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number87038
Device Catalogue Number87038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ULTRA ICE PLUS CATHETER
-
-