• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-95
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error code during procedure.Therefore, the customer switched to another device.After the procedure was completed, the customer tried to reproduce the error, but the error did not recur.There was no report of patient injury.
 
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-95 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.The temperature sensor have been calibrated as precaution.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the error reported by the customer remains unknown since the customer did not remember the code and no error could be reproduced during service activity.No additional information is available on this specific case and no further investigation is possible.Thus, the root cause could not be determined.No further complaint on this unit/ issue has been received up to date thus it is reasonable to assume that the error was temporary and did not recur.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key13079668
MDR Text Key283841131
Report Number9611109-2021-00724
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 12/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-95
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/24/2021
Initial Date FDA Received12/23/2021
Supplement Dates Manufacturer Received12/17/2021
01/24/2022
Supplement Dates FDA Received01/13/2022
02/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-