Model Number 72290129 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
Injury
|
Event Description
|
It was reported that, during a meniscus repair surgery, at the time the firstpass mini left was activated, it got trapped inside the joint cavity.When trying to remove the fragments, there was a risk of injury to the patient because it was trapped in the meniscus.All the fragments were removed.Surgery was resumed, after a non-significant delay, with a back-up device.No other complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference case (b)(4).
|
|
Manufacturer Narrative
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states no relevant supporting clinical information has been provided to assist with a clinical investigation.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.Per case details, all the fragments were removed from the patient.Should any additional clinical information be provided this complaint will be re-evaluated.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
|
Search Alerts/Recalls
|
|
|