Model Number MAJ-891 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2021 |
Event Type
malfunction
|
Event Description
|
Olympus medical systems corp.(omsc) was informed from the user that during the maintenance, the subject device was broken and fragment fell off.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
The subject device has been discarded at the customer site.Therefore the subject device was not returned to olympus.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Omsc could not reviewed the manufacture history (dhr) of the subject device because the lot number of the subject device was unknown.The exact cause of the reported event could not be conclusively determined.However, based on the reported information and past similar case, omsc presumed that the cause of the reported phenomenon was that the glue between the metal part and the plastic part deteriorated and peeled off due to physical stress/chemical stress and so on.If additional information becomes available, this report will be supplemented.
|
|
Search Alerts/Recalls
|