Model Number MAJ-891
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2021
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the maintenance, the subject device was broken and fragment fell off.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
The subject device has been discarded at the customer site.Therefore the subject device was not returned to olympus.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc could not reviewed the manufacture history (dhr) of the subject device because the lot number of the subject device was unknown.The exact cause of the reported event could not be conclusively determined.However, based on the reported information and past similar case, omsc presumed that the cause of the reported phenomenon was that the glue between the metal part and the plastic part deteriorated and peeled off due to physical stress/chemical stress and so on.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of Device
FORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13085606
MDR Text Key286755599
Report Number8010047-2021-16843
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-891
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/25/2021
Initial Date FDA Received12/26/2021
Supplement Dates Manufacturer Received02/07/2022
Supplement Dates FDA Received03/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-