MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was duplicated and the examination lamp of the subject device was not lit up due to a gas leak of the examination lamp.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Based upon the information from sorc, olympus medical systems corp.(omsc) surmised that the reported phenomenon occurred due to failure of the examination lamp (a gas leak of the examination lamp).The exact cause of a gas leak of the examination lamp could not be conclusively determined.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the unspecified timing, the emergency lamp of the subject device was always on.The event date was unknown.Other detailed information was not provided.There was no report of patient injury associated with the event.This device is an olympus asset.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13085755
MDR Text Key286109410
Report Number8010047-2021-16849
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown