RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2021 |
Event Type
malfunction
|
Event Description
|
Spo2 is not detecting readings correctly.The customer reported the device was in clinical use, however, there was no report of patient or user harm.The field service engineer (fse) evaluated the incident and advised caller to perform pvt flow testing as well as oxygen mix test to determine if flow sensor may be faulty.Caller stated that he has not performed testing, however he believes the flow sensor is faulty and requested part number.Further information is pending.
|
|
Manufacturer Narrative
|
Insufficient information is available to determine the resolution of the reported event.The customer was provided with the part id for flow sensor assembly; however, it could not be confirmed if the customer replaced the specified part.Multiple good faith efforts (gfe) were performed for further details regarding the event; however, no response was provided.No parts were returned for failure investigation.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|