It was reported that the stent migrated within 10 days.Through the attached photo it is confirmed that the stent was migrated.It was confirmed from the device history record that device had been manufactured with no significant issues and passed all the inspections successfully.Migration can occur in any company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.It is difficult to identify the exact root cause because the device was not returned, and it is difficult to reconstruct the situation at the time of procedure.However, based on the description "stent migrated within 10 days" and the migrated stent in the attached photo, it is assumed stent migration occurred due to severe pressure at the patient's lesion, peristalsis of organs, foreign substances such as food and other factors complexly, and was discharged from the body with feces.Through the user manual by taewoong, it is stated that "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: stent migration".This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
|