MAUDE Adverse Event Report: TAEWOONG MEDICAL CO.,LTD. NITI-S ENTERAL COLONIC COVERED STENT; COLONIC STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TAEWOONG MEDICAL CO.,LTD. NITI-S ENTERAL COLONIC COVERED STENT; COLONIC STENT Back to Search Results
Model Number CCT2212BA
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/19/2021
Event Type  malfunction  
Manufacturer Narrative
It was reported that the stent migrated within 10 days.Through the attached photo it is confirmed that the stent was migrated.It was confirmed from the device history record that device had been manufactured with no significant issues and passed all the inspections successfully.Migration can occur in any company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.It is difficult to identify the exact root cause because the device was not returned, and it is difficult to reconstruct the situation at the time of procedure.However, based on the description "stent migrated within 10 days" and the migrated stent in the attached photo, it is assumed stent migration occurred due to severe pressure at the patient's lesion, peristalsis of organs, foreign substances such as food and other factors complexly, and was discharged from the body with feces.Through the user manual by taewoong, it is stated that "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: stent migration".This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
 
Event Description
The stent was implanted on (b)(6) 2021 but when the physician checked on (b)(6) 2021 again, the stent disappeared, so that means this stent migrated within 10 days.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ENTERAL COLONIC COVERED STENT
Type of Device
COLONIC STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
lee
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
MDR Report Key13086344
MDR Text Key282772123
Report Number3003902943-2021-00052
Device Sequence Number1
Product Code MQR
Combination Product (y/n)Y
Reporter Country CodeTW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCCT2212BA
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/10/2021
Initial Date FDA Received12/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-