ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 600023011 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/20/2021 |
Event Type
malfunction
|
Event Description
|
During a supraventricular tachycardia radiofrequency procedure, a system error occurred which caused a delay.The error occurred multiple times while the system was running.Each time the error occurred the device was restarted, but the error would occur a few minutes later.The procedure was successfully completed with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
One ensite velocity¿ system velocity amplifier was received for evaluation.Visual inspection revealed the front panel ports, chassis, and labels were free of physical damage.The amplifier was powered on and the amplifier booted to a green led status.This indicated the amplifier passed the power on self-test (post).Communication was established and revision version 4.8 was observed.The field reported event was confirmed as review of error logs identified several voltage issues toward the catheter amplifier board at slot 11.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the root cause was isolated to the catheter amplifier boards at slot 11.
|
|