• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY ACROBAT¿ CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ACROBAT¿ CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number ACRO-35-450
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter; occupation: unknown.The investigation is on-going.A follow-up emdr will be provided within 30 days of submission of this report.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook acrobat¿ calibrated tip wire guide.It was reported that the physician detected that the coating of the wire guide peeled off during the procedure.The physician then changed to another of the same device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Investigation evaluation: our evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.Approximately 12.4cm to 27.3cm from the distal end is a section of bare core wire.Approximately 11.5cm to 12.4cm from the distal end, the wire guide covering has folded over itself.A section of the coating approximately 14.9cm long is frayed and hanging from the wire guide, and the coating is still attached at approximately 12.4cm from the distal end.Another section of coating approximately 4.9cm long is frayed and hanging from the wire guide, and the coating is still attached at approximately 22.4cm from the distal end.A conversation with production personnel from wire guides was had on 01/20/22.Pictures of the device were exchanged, and production management confirmed device was nonconforming due to coating damage caused during the scoring process in manufacturing.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the coating at the distal end was damaged which was due to the scoring process.This occurred due to operator error.Production management and the department team leads were notified of this occurrence.Additionally, a notification of operator related complaint form was provided to production management to make them aware of an operator related complaint.Prior to distribution, all acrobat¿ calibrated tip wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT¿ CALIBRATED TIP WIRE GUIDE
Type of Device
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key13088196
MDR Text Key288824565
Report Number1037905-2021-00715
Device Sequence Number1
Product Code OCY
UDI-Device Identifier10827002342665
UDI-Public(01)10827002342665(17)240723(10)W4498565
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberACRO-35-450
Device Lot NumberW4498565
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/28/2021
Initial Date FDA Received12/27/2021
Supplement Dates Manufacturer Received12/27/2021
Supplement Dates FDA Received01/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BOSTON SCIENTIFIC SPHINCTEROTOME; ENDOSCOPE, UNKNOWN MAKE OR MODEL; OLYMPUS HK260J ENDOSCOPE
Patient Age50 YR
Patient SexMale
Patient Weight75 KG
-
-