It was reported that, dressing was used to fix the picc tube, 5 days later the patient experienced scattered rashes and blisters with pruritus on the skin where the dressing of a iv3000 1 hand 10x12cm ctn 5 was applied.This adverse event was treated by oral antiallergic drugs, external application of trimetazidine cream on (b)(6) 2021.It is unknown current health status of patient.
|
H3, h6: the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review revealed a small number of similar instances in the last three years.There is nothing to indicate that this is outside of acceptable rates of occurrence.The device was used for treatment and was not returned for analysis.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.It was reported that 5 days after the dressing was applied the patient experienced scattered rashes and blisters.A clinical assessment determined that without clinically relevant patient-specific supporting documentation, a thorough medical investigation cannot be performed.The risk files mitigate the reported issue with no updates required.Ifu contains comprehensive instructions on skin preparation and application of dressing and also possible adverse reactions and wound suitability.A probable root cause for this issue is incorrect application of dressings or infrequent dressing changes.The users of the reported product are advised to consult the ifu, to prevent future occurrences of the reported issue.This guide provides comprehensive instructions of the operation, use and limitations of the device, including application and removal of dressings and skin preparation prior to use.This investigation is now complete, with no corrective actions required.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
|