MO016 MO-MOBERLY HOT PACK, LARGE, 6X9IN; PACK, HOT OR COLD, DISPOSABLE Back to Search Results
Model Number 11443-012B
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2021
Event Type  malfunction  
Event Description
Customer reported that the hot pack exploded onto the nurse's scrubs when the rn was squeezing the hot pack for activation.No injury reported.
 
Manufacturer Narrative
Device history review was completed on the reported lot v1p044.The lot was found to have been manufactured and released to predetermined specifications.No anomalies were found during review of the records.We received 1 sample for evaluation.The sample was already activated.The root cause was determined to be a void in the top seal resulting from incorrect machine setup.Personnel will be retrained on internal standard production methods and a warning label is being added to the label of the device that specifies that ¿when activating, hold away from patient.¿ we will continue to monitor complaint trends and utilize the information as part of continuous improvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOT PACK, LARGE, 6X9IN
Type of Device
PACK, HOT OR COLD, DISPOSABLE
Manufacturer (Section D)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer (Section G)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer Contact
patricia tucker
1500 waukegan rd
waukegan, IL 60085
8478874151
MDR Report Key13090756
MDR Text Key283246395
Report Number1423537-2021-00704
Device Sequence Number1
Product Code IMD
UDI-Device Identifier10885380169267
UDI-Public10885380169267
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number11443-012B
Device Catalogue Number11443-012B
Device Lot NumberV1P044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age45 YR
Patient SexFemale
-
-