• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL Back to Search Results
Model Number SE3690
Device Problems Device Displays Incorrect Message (2591); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 12/06/2021
Event Type  malfunction  
Event Description
According to the reporter, during a procedure, the unit had wrench light indicator error on.It was noted that there was a delay of the procedure.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a motor failure.It was reported that there was an adverse event without an identified device or use problem.A potentially related device issue could not be confirmed.The most likely cause could not be established from the information available.It was also reported that the device displayed an error of service required.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDVAC
Type of Device
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13095588
MDR Text Key282896349
Report Number1717344-2021-01690
Device Sequence Number1
Product Code FYD
UDI-Device Identifier10884524001555
UDI-Public10884524001555
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K980915
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSE3690
Device Catalogue NumberSE3690
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/13/2021
Initial Date FDA Received12/28/2021
Supplement Dates Manufacturer Received06/06/2022
Supplement Dates FDA Received06/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-