Catalog Number H938740
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Event Description
It was reported that two (2) 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags were leaking form the administration port.This issue was identified during compounding.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Initial reporter last name: (b)(6).Initial reporter city: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4, h6 and h10.H4: device manufactured between november 29, 2020 to november 30, 2020.H10: three (3) devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak at the spike port bonding area of all samples.The reported condition was verified.The cause was not determined; however, the most likely cause was due to inadequate, or lack of cyclohexanone being applied to the cap tubing when it was inserted to the spike port during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13096333
MDR Text Key282959245
Report Number1416980-2021-07554
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2023
Device Catalogue NumberH938740
Device Lot Number60269674
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2021
Initial Date FDA Received12/28/2021
Supplement Dates Manufacturer Received01/19/2022
Supplement Dates FDA Received01/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-