• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problems Crack (1135); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Event Description
A distributor in (b)(4) reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that base of a mr290v vented autofeed humidification chamber had a hole in the chamber base.There were no patient consequences.
 
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber has been received at fisher & paykel healthcare (b)(4).We will provide a follow up report upon completion of our investigation.
 
Event Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that base of a mr290v vented autofeed humidification chamber had a hole in the chamber base.There were no patient consequences.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was received at fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: visual inspection confirmed that multiple holes were found in the base of the chamber.Further analysis of the subject mr290v vented autofeed humification chamber confirmed the presence sodium chloride.Conclusion: the degradation of the aluminium baseplate is likely due to the presence of sodium chloride solution in the mr290v vented autofeed humidification chamber.Sodium salt is highly corrosive to aluminium.The presence of this solution could cause degradation of the aluminium plate over time.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.The subject chamber would have met the required specification prior to distribution.Our user instructions that accompany the mr290v vented autofeed humidification chamber states the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "use usp sterile water for inhalation or equivalent.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key13099753
MDR Text Key285226552
Report Number9611451-2021-01432
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number210228
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2021
Initial Date FDA Received12/28/2021
Supplement Dates Manufacturer Received03/01/2022
Supplement Dates FDA Received03/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-