Model Number CYF-V2
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2021
Event Type  malfunction  
Event Description
The customer reported that the distal end of her olympus videoscope tore off and had no eto cap.Additional details relating to the patient and the event have been requested, but no response has been received at this time.There was no patient harm or consequence reported as a result of this event.
 
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user report was confirmed, the distal end of the unit had torn away.Consequently, the device failed the scope leak test due to a leakage at the distal end.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the investigation, the cause of the distal end of the unit being torn away likely occurred from stress and/or user handling the device.The specific root cause could not be determined at this time.The following information is stated in the instructions for use: "chapter 3 preparation and inspection.3.2 inspection of the endoscope: 3.Inspect the external surface of the entire insertion tube for dents, bulges, swelling, peeling or other irregularities.4.Holding the insertion tube gently with one hand, carefully run your fingertips over the entire length of the insertion tube in both directions.Inspect for any protruding objects or other irregularities.Also confirm that the insertion tube is not abnormally rigid." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE "CYF-V2", UK VERSION
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13101505
MDR Text Key287284892
Report Number8010047-2021-16924
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339431
UDI-Public04953170339431
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/28/2021
Supplement Dates Manufacturer Received01/14/2022
Supplement Dates FDA Received01/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-