Model Number MR290V
Device Problems Leak/Splash (1354); Overfill (2404)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber has been received at fisher & paykel healthcare (b)(6).We will provide a follow up report upon completion of our investigation.
 
Event Description
A distributor in (b)(6) reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that base of a mr290v vented autofeed humidification chamber was overfilled during use.There were no patient consequences.
 
Event Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that base of a mr290v vented autofeed humidification chamber was overfilled during use.There were no patient consequences.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected.Results: visual inspection of the mr290v chamber confirmed that there were no notable damages.There were no leaks identified when the device was connected to a water source.Conclusion: we are unable to confirm the reported event as there was no fault found with the returned mr290v chamber.Every mr290 chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure" - "ensure there is water supply connected to the chamber and that water is present within the chamber" - "do not use the chamber if the seals are not intact when received, or if it has been dropped.''.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key13102122
MDR Text Key285615672
Report Number9611451-2021-01507
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number200711
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/28/2022
Initial Date FDA Received12/28/2021
Supplement Dates Manufacturer Received01/28/2022
Supplement Dates FDA Received01/31/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
F&P MR850 RESPIRATORY HUMIDIFIER; RT104 BREATHING CIRCUIT
-
-