• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE Back to Search Results
Model Number M0063903200
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2021
Event Type  malfunction  
Manufacturer Narrative
The returned stone cone was analyzed, and a visual evaluation found the coil exposed and tangled.The green coating of the coil was damaged such that the inner coil wire is exposed.Additionally, the outer blue sheath had a slight tear at the distal end.Due to the defects, functional testing found the coil could not be successfully opened or closed.Therefore, the reported event is confirmed.Based on all available information, it is highly unlikely that the defects described were present upon opening of the packaging.It is most likely that they were caused using excessive force or manipulation during testing.The instructions for use (ifu) warns the user, prior to use, ensure that the coil is working properly by advancing the sheath over the coil to the positive stop and then retracting the sheath to open the coil.The sheath of the device should be straight during testing.Therefore, the most probable root cause is unintended use error caused or contributed to event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that a stone cone nitinol retrieval coil was to be used in the ureteral calculus ureteroscopic holmium laser lithotripsy procedure performed on (b)(6) 2021.During preparation, the stone cone was tested and found the device failed to form a coil.Another stone cone was used to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr-reportable event based on the investigation results that the green coating of the coil was damaged and inner coil wire is exposed, and outer blue sheath had a slight tear at the distal end.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONE CONE
Type of Device
DISLODGER, STONE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
LAKE REGION MEDICAL
31c butterfield trail
el paso TX 79906
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13110355
MDR Text Key284717501
Report Number3005099803-2021-07928
Device Sequence Number1
Product Code FGO
UDI-Device Identifier08714729430223
UDI-Public08714729430223
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2023
Device Model NumberM0063903200
Device Catalogue Number390-320
Device Lot Number0005661511
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight75 KG
-
-