BAXTER HEALTHCARE CORPORATION EM 2400, MAIN MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING
|
Back to Search Results |
|
Catalog Number 2400M |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the power cord of an exactamix em 2400, main module was damaged.The damaged power cord was further described to have exposed wires.It was not reported where in the process the event occurred.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.During visual inspection it was observed that the faulty power cord was not provided.The device was functionally tested with a replacement power cord and performed according to product specifications.The reported condition was not verified.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.To correct the condition, the power cord was replaced.Should additional relevant information become available, a supplemental report will be submitted.
|
|