Model Number V60
Device Problem Inadequate User Interface (2958)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Event Description
The customer called into technical support (ts) reporting that the device has error 1105 ventilator led alarm failed message.It is unknown if the unit was in patient use at the time of the reported issue.No patient or user harm was reported.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed reported problem.Also, provided part number for pm board.Further information is pending.
 
Manufacturer Narrative
The customer replaced the power switch overlay to resolve the reported issue.The device passed required performance verification tests per philips¿s standards.
 
Manufacturer Narrative
H11 patient involvement: the customer reported that there was no patient involvement at the time the issue was discovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key13114332
MDR Text Key283110622
Report Number2031642-2021-05911
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2021
Initial Date FDA Received12/29/2021
Supplement Dates Manufacturer Received01/06/2022
03/16/2022
Supplement Dates FDA Received01/31/2022
04/11/2022
Date Device Manufactured04/07/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-