• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON CARIBE LTD. BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON CARIBE LTD. BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING Back to Search Results
Model Number 442023
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that while testing with the bd bactec¿ plus aerobic/f culture vials (plastic), one false positive result was obtained.Confirmatory subculture testing was performed.Erroneous results were not reported out, and there was no report of patient impact.The following information was provided by the initial reporter: 1 false positive bottle on 442023 lot 1245128 due to poor sensor adhesion.
 
Event Description
It was reported that while testing with the bd bactec¿ plus aerobic/f culture vials (plastic), one false positive result was obtained.Confirmatory subculture testing was performed.Erroneous results were not reported out, and there was no report of patient impact.The following information was provided by the initial reporter: 1 false positive bottle on 442023 lot 1245128 due to poor sensor adhesion.
 
Manufacturer Narrative
Investigation summary: customer reported a sensor adhesion defect.Photos were not received.Retention samples could not be visually inspected/ tested for reported defect.Batch history record review was used for this investigation.Batch/sensor history records were reviewed, and all testing were within specification for product release.A complaint history review was conducted and only the current complaint was found relating to the incident lot number and the ¿as reported¿ defect code.Sensor adhesion scrape test is performed to each sensor batch as part of release criteria.Complaint is unconfirmed.The vision system is challenged prior each lot.Product insert warnings and precautions sections states that prior to use, each vial should be examined for evidence of contamination such as cloudiness, bulging or depresses septum, or leakage.Vials showing evidence of contamination should not be used.Also prior to use, the user should examine the vial for evidence of damage or deterioration.Vials displaying turbidity, contamination, or discoloration (darkening) should not be used.Corrective actions preventive actions (capa) 2882676 was initiated to further investigate these types of complaints and determine any appropriate actions to reduce their occurrence.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigational process.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13114399
MDR Text Key285493036
Report Number2647876-2021-00380
Device Sequence Number1
Product Code MDB
UDI-Device Identifier00382904420239
UDI-Public00382904420239
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113558
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2022
Device Model Number442023
Device Catalogue Number442023
Device Lot Number1245128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/29/2021
Supplement Dates Manufacturer Received02/04/2022
Supplement Dates FDA Received02/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-