Catalog Number 2C4702K |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a small volume folfusor leaked at the white tip connected to a needle.This was observed during patient infusion.A transparent, occlusive dressing was applied as a patch.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received containing an unspecified volume of fluid in the bladder.Visual inspection using the naked eye noted a red luer cap attached to the distal luer ("white tip").The red luer cap is not a baxter product.A leak test was performed and no evidence of leak was observed at the "white tip" (distal luer) or at other parts of he folfusor device.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|