RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Inadequate User Interface (2958)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Event Description
|
It was reported to philips by a customer that the unit's bottom half of touchscreen not working properly.Based upon the information provided, it is unknown if the unit was in use on a patient at the time of the reported event, however, no patient harm or injury was reported.The device was evaluated remotely by a philips remote service engineer (rse).Upon further inspection and review of the device, the biomed stated that touchscreen had dead spots even after calibration passes.The rse advised the biomed to replace the touchscreen and provided the part number.It is unknown if any parts or repair has been conducted in relation to the alleged complaint.Attempts to obtain further information are currently pending.
|
|
Manufacturer Narrative
|
The customer replaced the touchscreen to resolve the reported issue.The unit was functionally tested and successfully passed all testing.The unit was returned to service.No other anomaly was reported.
|
|