MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number CXT261412
Device Problem Insufficient Information (3190)
Patient Problem Aneurysm (1708)
Event Date 12/22/2021
Event Type  Injury  
Event Description
On (b)(6) 2021 the patient underwent treatment for an abdominal aortic aneurysm and was implanted with gore ® excluder ® conformable aaa endoprosthesis (cxt261412) and with gore ® excluder ® aaa endoprostheses (plc14100, plc201200, and plc121400).The plc 121400 was extended into the ipsilateral limb on the right side of the patient.A 3cm of overlap was observed between the cxt261412 and the plc121400 at the time of the procedure.A follow up ct scan was performed on an unspecified date and it was observed that the there was no overlap between cxt261412 and the plc121400 endoprostheses.The plc121400 had migrated to the end of the ipsilateral limb on the cxt261412.A type iii endoleak was observed.On (b)(6) 2021 a reintervention was performed in which a plc 161400 was implanted to treat the type iii endoleak.The patient tolerated the procedure and the endoleak was resolved.
 
Manufacturer Narrative
H6 code 213 - no device problem found: a review of manufacturing records verified that the lot involved in this complaint met all pre-release specifications.
 
Manufacturer Narrative
G3/g4, pma/510k number incorrectly entered previously and changed to p200030.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Type of Device
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
breanna cox
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13114635
MDR Text Key288184952
Report Number3007284313-2021-01732
Device Sequence Number1
Product Code MIH
UDI-Device Identifier00733132651030
UDI-Public00733132651030
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P200030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/22/2023
Device Model NumberCXT261412
Device Catalogue NumberCXT261412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2021
Initial Date FDA Received12/29/2021
Supplement Dates Manufacturer Received12/22/2021
03/16/2022
Supplement Dates FDA Received01/04/2022
03/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Other;
Patient Age71 YR
Patient SexMale
-
-