The meter and test strip vial were provided for investigation.However, the returned test strip vial had no remaining test strips.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling: "coaguchek method uses human recombinant thromboplastin.Therefore, the comparability to tests using other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastin types.However, those higher differences between thromboplastins of different (rabbit, bovine) origin are not an issue specific for coaguchek assays.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared with several other (rabbit, bovine) laboratory methods." the investigation is ongoing.Occupation is patient/consumer.
|
The customer's returned meter was tested with retention test strips and retention high level control sample.Testing results (qc range = 2.5 - 3.7 inr): qc 1: 3.2 inr, qc 2: 3.1 inr, qc 3: 3.1 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.
|