Model Number 865351
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2021
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reports that the telemetry pack experienced a speaker malfunction and no noise.The device was not in use on a patient at the time of event, there was no patient involvement.
 
Event Description
The customer reports that the telemetry pack experienced a speaker malfunction and no noise.The device was not in use on a patient at the time of event, there was no patient involvement.
 
Manufacturer Narrative
The device was returned for evaluation at a philips authorized repair facility.Evaluation of the device determined the speaker produced audible sound.There was a speaker malfunction message present due to the use of a third party speaker.Evaluation of the device determined that the speaker did produce audible sound.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 2.4 GHZ SMART HOPPING
Type of Device
MX40 2.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
3000 minuteman road
andover MA 01810
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
3000 minuteman road
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key13118592
MDR Text Key283142815
Report Number1218950-2021-11224
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082243
UDI-Public00884838082243
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865351
Device Catalogue Number865351
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/09/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received04/15/2022
01/23/2023
Supplement Dates FDA Received04/18/2022
02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-