• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TRACHEOSTOMY TUBE AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TRACHEOSTOMY TUBE AND TUBE CUFF Back to Search Results
Model Number 101/860/080CZ
Device Problem Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2021
Event Type  malfunction  
Manufacturer Narrative
One unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Due to fact that cuff leak was not observed prior use it is the most probable that reported failure occurred during tracheostomy procedure or during use.Dhr review was done, no issues related to the original complaint were found.
 
Event Description
It was reported that when the customer inserted the product in the patient and inflated the cuff, its air pressure was unable to be maintained.No patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
TRACHEOSTOMY TUBE AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
david halverson
6000 nathan lane north
7-15-26 fukushima
minneapolis, MN 55442
MDR Report Key13118740
MDR Text Key283682478
Report Number3012307300-2021-13509
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076158
UDI-Public15019517076158
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/080CZ
Device Lot Number4079448
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2021
Initial Date FDA Received12/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-