ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA25-120/I20-40
Device Problems Material Puncture/Hole (1504); Material Integrity Problem (2978)
Patient Problem Failure of Implant (1924)
Event Date 12/22/2021
Event Type  Injury  
Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with strata.
 
Event Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx bifurcated stent graft.Approximately eight and a half (8.5) years post initial procedure, a type 3b endoleak was identified.Reintervention is planned; however, has not yet taken place.Patient status was not provided.
 
Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows type iiib endoleak is unconfirmed.This is not consistent with the reported adverse event/incident.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported to be pending a reintervention.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with strata.G3: awareness date ¿ updated.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Type of Device
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13119362
MDR Text Key282984958
Report Number2031527-2021-00512
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2013
Device Model NumberBA25-120/I20-40
Device Lot Number1111204-020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received02/13/2022
Supplement Dates FDA Received03/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/05/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0006-2019
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age64 YR
Patient SexMale
-
-