MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
|
Back to Search Results |
|
Model Number NEU_INS_STIMULATOR |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Pain (1994)
|
Event Date 12/20/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: th90p02, serial#: unknown, product type: mobile platform.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that representative the patient received a smart programmer after implant in (b)(6) 2021.She had issues with making connection.Also the nurse did not succeed in making connection with the ipg 3058.It was noted that the patient requested to have the ipg implanted deeper, due to pain issues in the past.The connection with icon was easy.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a manufacturer representative (rep).The rep reported that they were referring to the battery consumption of the smart programmer.They checked device maintenance in the settings menu.They found out the battery of the smart programmer was heavily used and only had 10% left.They think this had to do with all the apps that were working in the background.They pressed the button optimize etc.Now the screen showed that all is well, 100%.They thought that the bad maintenance/high consumption of the battery of the smart programmer could have been the cause of the connection issues.
|
|
Manufacturer Narrative
|
H6 updated to reflect the correct decision.Review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information indicated that the smart programmer was not returned and still in use.Hcp was notified of the event.
|
|