• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER Back to Search Results
Model Number 518-019
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 12/08/2021
Event Type  Injury  
Manufacturer Narrative
Patient's weight unk.The device was discarded, thus no investigation could be completed.
 
Event Description
A lead extraction procedure commenced to remove a right ventricular (rv) and a right atrial (ra) lead due to cied system/pocket infection.Spectranetics lead locking devices (llds) were inserted into each lead to provide traction.A spectranetics 16f glidelight laser sheath was calibrated and used in the patient for a few seconds.An area on the device was observed to be knotted/kinked (kink could have occurred during handling of the device during calibration).The device's laser fibers became apparent through that knotted/kinked area with just a few seconds of lasing (mdr #1721279-2021-00256).The device was discarded; another glidelight device was opened and used.However, after several attempts with different tools including snaring from the groin, the rv and ra leads, with llds present within each lead, were cut and capped and remained in the patient (mdr #1721279-2021-00258 lld in ra lead).The physician did not attempt to unlock the llds from the leads prior to cutting and capping.The patient survived the procedure.This report captures the lld which remained in the rv lead when it was cut/capped by the physician.There was no alleged malfunction of the llds used in the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key13120221
MDR Text Key283086309
Report Number1721279-2021-00257
Device Sequence Number1
Product Code DRB
UDI-Device Identifier00813132023014
UDI-Public(01)00813132023014(17)230716(10)FLC21G14A
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990713
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/16/2023
Device Model Number518-019
Device Catalogue Number518-019
Device Lot NumberFLC21G14A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/08/2021
Initial Date FDA Received12/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BOSTON SCIENTIFIC 0185 RV ICD LEAD.; BOSTON SCIENTIFIC 4470 RA PACING LEAD.; FEMORAL SNARE MANUFACTURER UNK.; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM.; SPECTRANETICS GLIDELIGHT LASER SHEATHS.; SPECTRANETICS LEAD LOCKING DEVICE IN RA LEAD.; SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH.
Patient Outcome(s) Other;
Patient Age50 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-