SMITH & NEPHEW, INC. BIOSURE RATCHET DRIVER; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72201888
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during surgery using a biosure ratchet driver, it was noticed that the product key was broken.The procedure was completed without delay, but it is unknown how.No further complications were reported.
 
Manufacturer Narrative
H3, h6: ¿the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.It was determined the device did not contribute to the reported event.A complaint history review concluded this was a repeat issue.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the customer provided image found a picture of the distal end of the device.The tip of the device is broken.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.¿ corrected data: b5: event description h1: type of reportable event h6: health effect - impact code.
 
Event Description
It was reported that, during surgery using a biosure ratchet driver, it was noticed that the product key was broken.It is unknown how the procedure was completed.No delay and no further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE RATCHET DRIVER
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13120340
MDR Text Key283084287
Report Number1219602-2021-02813
Device Sequence Number1
Product Code NBH
UDI-Device Identifier03596010615442
UDI-Public03596010615442
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72201888
Device Catalogue Number72201888
Device Lot Number50563816
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/15/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received02/14/2022
Supplement Dates FDA Received02/15/2022
Date Device Manufactured09/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-