Model Number 1003327 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during device preparation of the indeflator, when negative was pulled, air entered the indeflator.The indeflator was not used.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.One nonconforming material record/exception was generated for the finished good lot to investigate the reported issue in accordance with internal operating procedures.The investigation determined the reported difficulties appear to be a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.D9, h3: device not available for evaluation.
|
|
Manufacturer Narrative
|
On feb 25th, 2022 abbott vascular decided to initiate a voluntary field action for this product.Abbott vascular submitted medwatch # 2024168-2022-01831 on march 4, 2022 with notification of the voluntary recall in h7, (remedial action initiated).Corrective action has been initiated per site operating procedures.The product will continue to be trended.This action is being taken due to an increase in the complaint trend for reported device code leak/splash, and loose or intermittent connection.20/30 indeflators are at an increased risk of leaking due to a gap in the hose snap fitting.Stopcocks are at an increased risk of leaking due to a higher tendency for loose connections when not connected properly.
|
|