Model Number MS20401 |
Device Problem
No Audible Prompt/Feedback (2282)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 12/01/2021 |
Event Type
Death
|
Event Description
|
It was reported by the site biomedical engineer that there was an ecg leads off condition for approximately 45 minutes and the m540 monitor did not alarm and the involved patient died.At this time the exact date of the event was unknown, only that it was 2-3 weeks prior to (b)(6) 2021.[date of death estimated as (b)(6) 2021] the site biomed stated they were not notified when the event occurred, the monitor was not taken out of use and the event was not locked at the infinity central station (ics) ics.Therefore, the event logs from the m540 and the ics may have been overwritten.The involved device and others were checked out and it was confirmed that they function properly and alarmed for the ecg leads off situation.It was noted that the ecg leads off alarm priority was set at low (default) and the alarm volume setting was 20%.Additionally, it was noted that the ics is not monitored 100% of the time and that staff may have selected ¿privacy mode¿ on the bedside monitor.
|
|
Manufacturer Narrative
|
A follow-up report will be submitted upon completion of this investigation.
|
|
Manufacturer Narrative
|
As previously stated, the site biomed tested the device and confirmed that the device was properly alarming for an ecg leads off condition at the time of testing.It was found that the leads off alarm was configured as a low priority alarm and the device alarm volume was set to 20%.After the event, the site has updated their device settings to make alarms more easily identifiable.After further communication with the site, it was also determined that the ics is not monitored at all times and it was possible that "privacy mode" had been turned on, both of which could prevent alarms from being noticed by the user.Additionally, it was noted that the patient was also connected to a ventilator during the event.The ventilator mask had either fallen off or was pulled off by the patient and the staff indicated that the ventilator also did not alarm.The ventilator logs were reviewed and it was confirmed that the ventilator did issue alarms.The date of the event was clarified as on (b)(6) 2021.The involved m540 and ics log files provided were reviewed by draeger engineering.It was determined that the log files from the date of the reported event were overwritten and therefore could not aid in this investigation.A draeger technician evaluated the device with no malfunctions identified.Therefore, we are unable to verify the reported failure to alarm and root cause cannot be definitively determined.H3 other text: a draeger technician evaluated the device with no malfunctions identified.
|
|
Event Description
|
It was reported by the site biomedical engineer that there was an ecg leads off condition for approximately 45 minutes and the m540 monitor did not alarm and the involved patient died.At this time the exact date of the event was unknown, only that it was 2-3 weeks prior to on (b)(6).[date of death estimated as on (b)(6) 2021].The site biomed stated they were not notified when the event occurred, the monitor was not taken out of use and the event was not locked at the infinity central station (ics) ics.Therefore, the event logs from the m540 and the ics may have been overwritten.The involved device and others were checked out and it was confirmed that they function properly and alarmed for the ecg leads off situation.It was noted that the ecg leads off alarm priority was set at low (default) and the alarm volume setting was 20%.Additionally, it was noted that the ics is not monitored 100% of the time and that staff may have selected ¿privacy mode¿ on the bedside monitor.
|
|
Event Description
|
It was reported by the site biomedical engineer that there was an ecg leads off condition for approximately 45 minutes and the m540 monitor did not alarm and the involved patient died.At this time the exact date of the event was unknown, only that it was 2-3 weeks prior to (b)(6) 2023.[date of death estimated as (b)(6) 2021].The site biomed stated they were not notified when the event occurred, the monitor was not taken out of use and the event was not locked at the infinity central station (ics) ics.Therefore, the event logs from the m540 and the ics may have been overwritten.The involved device and others were checked out and it was confirmed that they function properly and alarmed for the ecg leads off situation.It was noted that the ecg leads off alarm priority was set at low (default) and the alarm volume setting was 20%.Additionally, it was noted that the ics is not monitored 100% of the time and that staff may have selected ¿privacy mode¿ on the bedside monitor.
|
|
Manufacturer Narrative
|
Due to a technical issue with our internal emdr system we submitted for the form fda 3500a an incorrect value for the field h3.- not returned to manufacturer.H3 other text : a draeger technician evaluated the device with no malfunctions identified.
|
|
Search Alerts/Recalls
|