• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4705K
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2021
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the bladder of a small volume folfusor ruptured during filling with glucose.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6 and h10: h10: the device was received for evaluation.During visual inspection, the bladder was observed ruptured.The external and internal surfaces of the bladder and the rupture line were microscopically examined for evidence of markings, abrasions, gouges, holes, or embedded particulate matter along with any surface defects on the stress-member for any signs of abnormality which may have caused the device to rupture.No signs of abnormality were found on the ruptured bladder.The reported condition was verified.The cause of the condition could not be determined, however, the potential for bladder ruptures exists as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13122101
MDR Text Key283094408
Report Number1416980-2021-07624
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412081588
UDI-Public(01)00085412081588
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C4705K
Device Lot Number21A010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/24/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/08/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received01/11/2022
Supplement Dates FDA Received01/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-