(b)(6) 2022 device explanted.The icd was returned for analysis.Before analyzing the icd, the manufacturing process for this device was re-investigated and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.Particularly the final acceptance test proved the device functions to be as specified.Incoming visual inspection revealed electro-cautery and arc-over marks on the icd housing.Upon receipt, the device could not be interrogated.Therefore, the device was opened to inspect the inner assembly and to check the functionality of the electronic module.Analysis of the electronic module showed that an output stage of a high voltage circuit and several other electric components had been damaged.These damages led to an elevated current consumption, which depleted the battery.As a result of the battery depletion and the damages of the electronic module the device could not be interrogated.The battery was sent to the manufacturer for further analysis.In depth inspection of the battery showed no anomalies.Based on the damage symptoms of the electronic module, the damages were most probably caused by an external defibrillation.The exact external root cause is not determinable based on the available information.However, a defective icd can be excluded.
|