Model Number CV-180
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2021
Event Type  malfunction  
Manufacturer Narrative
Olympus technical assistance center (tac) support called customer to troubleshoot the device.Customer reported that there was no connection from the pc to the cv-180 device.Tac advised customer to reset the cv-180 firmware and select cvp under user settings, and the issue was resolved.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
A user facility returned the olympus evis exera ii video system center due to the device not connecting to the personal computer (pc) and not displaying any image.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The subject device was manufactured on a date that was not identified.The device history record was unable to be reviewed for this device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The case is not due to design.Based on the results of the legal manufacturer's investigation, it was presumed that the indicated phenomenon was caused by a firmware defect, but the cause of the firmware defect could not be identified.There was a problem with the cvp selected in the user settings that caused the error to the firmware resetting.The issue was resolved.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13122624
MDR Text Key284715267
Report Number8010047-2021-16998
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received03/01/2022
Supplement Dates FDA Received03/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-