RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported to philips that a noise was heard mixed with resonance.The device was reported to have been in patient use.There was no harm, delay in therapy, or medical intervention reported.The philips field service engineer (fse) confirmed and duplicated the reported issue.The vibration dampening ring needed to be repositioned.Following the evaluation of the device, the fse repositioned the vibration dampening ring to resolve the problem.The unit was then functionally tested and successfully passed specified tests.No other abnormality was observed.
|
|