Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported to philips that a noise was heard mixed with resonance.The device was reported to have been in patient use.There was no harm, delay in therapy, or medical intervention reported.The philips field service engineer (fse) confirmed and duplicated the reported issue.The vibration dampening ring needed to be repositioned.Following the evaluation of the device, the fse repositioned the vibration dampening ring to resolve the problem.The unit was then functionally tested and successfully passed specified tests.No other abnormality was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key13122776
MDR Text Key286596781
Report Number2031642-2021-05917
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838033832
UDI-Public00884838033832
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/06/2021
Initial Date FDA Received12/30/2021
Date Device Manufactured09/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse