• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. KNEE SCORPION; ACCESSORIES, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. KNEE SCORPION; ACCESSORIES, ARTHROSCOPIC Back to Search Results
Model Number KNEE SCORPION
Device Problem Break (1069)
Patient Problem Failure of Implant (1924)
Event Date 12/15/2021
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On (b)(6) 2021, it was reported by a sales representative via email that on an ar-12990 knee scorpion top part fell off.This was discovered during a procedure.Additional information received 12/22/2021: the top jaw of the knee scorpion broke off inside the patient during a root repair procedure.Case was completed by removing broken fragments with a kingfisher and using another knee scorpion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE SCORPION
Type of Device
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13154822
MDR Text Key286248883
Report Number1220246-2022-04247
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00888867196322
UDI-Public00888867196322
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 01/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKNEE SCORPION
Device Catalogue NumberAR-12990
Device Lot Number82845
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/15/2021
Initial Date FDA Received01/04/2022
Date Device Manufactured08/21/2017
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-