MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number PXMK11076
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that about 1mm size of unknown white-floating material was found inside the pressure monitoring kit line in an opened pouch during priming.Patient demographic information is unknown.There were no patient complications reported.
 
Manufacturer Narrative
The reported event of unknown white-floating material inside the line was confirmed.One white material was observed at flow path of male connector, which was connected to female luer of merit flush device.The material was approximately 1 x 1 mm in size.The material moved during continuous flushing, but the material was stuck at iv side of merit flush device and stayed at the location after 5 minutes of continuous flushing.Further evaluation regarding white material is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key13161104
MDR Text Key283236879
Report Number2015691-2022-03219
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K925638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/15/2023
Device Model NumberPXMK11076
Device Lot NumberPF1408MT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2021
Initial Date FDA Received01/04/2022
Supplement Dates Manufacturer Received02/03/2022
Supplement Dates FDA Received02/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-