• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Contamination /Decontamination Problem (2895); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed by the customer that the exhaust valve did not close.There was no report of patient injury associated with the event.The event date was unknown.
 
Manufacturer Narrative
The subject device was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon could not be duplicated.It was also found that black liquid was in the pipeline due to insufficient reprocessing.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The manufacturing record was reviewed and found no irregularities.Omsc re-evaluated the reported event that the exhaust valve did not close, and judged that it was not reportable.According to the inspection result by sorc, black liquid was in the pipeline, but no other abnormalities were found.The exact cause of black liquid in the pipeline could not be conclusively determined.
 
Event Description
This report was submitted for black liquid in the pipeline due to insufficient reprocessing found during device evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13166615
MDR Text Key285433710
Report Number8010047-2022-00886
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2021
Initial Date FDA Received01/05/2022
Supplement Dates Manufacturer Received01/12/2022
04/12/2022
Supplement Dates FDA Received02/02/2022
05/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-