Catalog Number 60-01-00
Device Problem Decreased Pump Speed (1500)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The centrifugal pump system with tubing clamp 60-01-00 is not distributed in the usa and it is similar to centrifugal pump system with tubing clamp 60-02-15, which is distributed in the usa (510(k) number: k032213).Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to the motor control board which was replaced to solve the reported failure.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp went from 2000 rpm down to 1000 rpm during service.There was no report of patient injury.
 
Manufacturer Narrative
H.10: the root cause of the reported issue is a defective motor control board.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key13168337
MDR Text Key285696490
Report Number9611109-2022-00003
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number60-01-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/08/2021
Initial Date FDA Received01/05/2022
Supplement Dates Manufacturer Received02/07/2022
Supplement Dates FDA Received03/07/2022
Date Device Manufactured11/14/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-