• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 ANATOMIC GLENOID L 29; SHOULDER IMPLANT - GLENOID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 ANATOMIC GLENOID L 29; SHOULDER IMPLANT - GLENOID Back to Search Results
Model Number 5100-29-000
Device Problems Fracture (1260); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
One tab on the glenoid inserter tip, broke during impaction.No adverse affects on patient.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: h6 (device).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was not returned for evaluation.The investigation confirmed that the inserter tip, sold assembled to the glenoid implant, was broken.The provided photograph does not represent the finished good implant device.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANATOMIC GLENOID L 29
Type of Device
SHOULDER IMPLANT - GLENOID
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13172648
MDR Text Key285687438
Report Number1818910-2022-00264
Device Sequence Number1
Product Code MBF
UDI-Device Identifier10603295538134
UDI-Public10603295538134
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5100-29-000
Device Catalogue Number510029000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2021
Initial Date FDA Received01/05/2022
Supplement Dates Manufacturer Received02/18/2022
02/25/2022
Supplement Dates FDA Received02/23/2022
02/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-