• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX TI 2.8 DURABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX TI 2.8 DURABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72200762
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2021
Event Type  malfunction  
Event Description
It was reported that during an arthroscopy, the shaft of twinfix was found to be engraved with four laser marks, instead of two laser marks.It is unknown how the procedure was completed or if there was a delay.No patient complications were reported.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
H3, h6: the reported device was received for evaluation.It was determined the device contributed to the reported event.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the shaft drawing found that there are two axial laser marks 180° apart that correspond to the center of the opposite flats on the hex.A visual inspection of the returned device found that it is not in its original packaging.The device has been deployed, and the anchor and sutures were not returned.There is debris on the device.There are four laser vertical laser markings on the shaft of the device.Two markings on each side, approximately 180° apart.The markings are worn and scuffed.The complaint was confirmed, and the root cause was associated with manufacturing.A quality alert has been issued in response to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX TI 2.8 DURABRAID
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13175718
MDR Text Key283328761
Report Number1219602-2022-00035
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010584458
UDI-Public03596010584458
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200762
Device Catalogue Number72200762
Device Lot Number2067740
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2021
Initial Date FDA Received01/06/2022
Supplement Dates Manufacturer Received04/25/2022
Supplement Dates FDA Received04/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-