The patient expired a day after the device failed to alarm.Biomed tested the system afterwards and everything worked fine.Technical support (ts) had the customer print out the investigation documents.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Concomitant medical device: the following device was used in conjunction with the cns: co2 monitor: model #: ni, serial #: ni, device manufacturer data: ni, unique identifier (udi) #: ni, returned to nihon kohden: ni.
|
Details of complaint: the customer reported that the central nurse's station (cns) did not alarm for apnea.According to the customer, the patient was being monitored normally, but when the patient went into respiratory arrest, the cns did not alarm.The patient expired a day after the device failed to alarm.Biomed tested the system afterwards and everything worked fine.Technical support (ts) had the customer print out the investigation documents.Investigation summary: based on the information provided by nka clinical in ticket (b)(4), the bedside monitor did not alarm for respiration due to having a co2 monitor being connected to the bedside though it was not being used to monitor the patient.The central monitoring station (cns) did not alarm for respiration due to the customer not having the parameter for respiration from impedance (rr imp) selected for monitoring on the cns.The printouts provided by the customer is also evidence which shows that rr imp was not being monitored at the time.The operator's manual for pu-6801ra, page 9.21, has information for selecting parameters for review.The root cause for this issue is use error.As this issue has an overall risk score of medium, a capa is not required per corrective action and preventive action process, sop07-003." manufacturer references # 300277611-127934 follow up 001.
|